چیزی ک تو سرویس منوال پیدا کردم
C6000 Broken fixing heater wire
In fixing warm-up, the time to reach
50°C/122°F exceeds 13.5 s, the time to
reach 100°C/212°F exceeds 10 s, the
time to reach the primary stabilization
exceeds 10 s or the time to reach the
secondary stabilization exceeds 24 s.
Poor contact in the
thermistor connector
terminals.
Reinsert the connector. Also check for continuity
within the connector cable. If none,
remedy or replace the cable.
Fixing thermistor
installed incorrectly.
Check and reinstall if necessary.
Fixing thermostat
triggered.
Check for continuity. If none, replace the fixing
thermostat.
Code Contents
Remarks
Causes Check procedures/corrective measures
Downloaded from
میشه گفت مشکل از فیوزینگ و نداشتن حرارت کافی باشه ک باید کابل ها و ترمیستور چک بشه . امیدوارم نتیجه بده